Cardiac Science Notifies AED Customers of Nationwide Voluntary Medical Device CorrectionBOTHELL, Wash., Nov 13, 2009 /PRNewswire-FirstCall via COMTEX News Network/
-- Cardiac
Science Corporation (Nasdaq: CSCX) is initiating a voluntary field
correction after it was determined certain automated external defibrillators
(AEDs) may experience a rare product issue in which the AED may not be able to
deliver therapy during a resuscitation attempt. Device failure may affect
resuscitation of the patient, which could lead to serious adverse events or
death. These AEDs have electronic components which may fail and the failure may
not be detected by the device's periodic self-tests. The affected models include
the Powerheart 9300A, 9300C, 9300D, 9300E, 9300P, 9390A, 9390E, and CardioVive
92531, 92532, and 92533 devices. Read more about the Cardiac Science Medical Device Correction
|
|
AEDUniverse.com is your online source for Cardiac Science
AED product manuals, troubleshooting instructions, recall
information, warranty information, product registration and help for
your Cardiac Science Automated External Defibrillators or AED's.
|
|
|